Overview, Specifications – RKI Instruments 35-3001-04 User Manual

Page 5

Advertising
background image

35-3001-04 Carbon Monoxide Sample-Draw Detector • 1

Overview

This operator’s manual describes the 35-3001-04 carbon monoxide sample-draw detector. This
manual also describes how to install, start up, maintain, and calibrate the sample-draw detector
when using it with a gas monitoring controller. A parts list at the end of this manual lists
replacement parts and accessories for the sample-draw detector.

Specifications

Table 1 lists specifications for the carbon monoxide sample-draw detector. See the controller
Operator’s Manual for information specific to the controller.

WARNING:

When using the 35-3001-04, you must follow the instructions and warnings in
this manual to assure proper and safe operation of the 35-3001-04 and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 35-3001-04 as described in this manual.

Table 1: Specifications

Target Gas and Detection Range

Carbon Monoxide: 0 - 300 ppm

Input Power

24 VDC Nominal (18.5 VDC - 30 VDC)

Construction (housing)

Fiberglass/polyester (NEMA 4X)

Dimensions

8.5 in. H x 6.5 in. W x 4.25 in. D

Weight

4.5 lbs.

Sampling Method

Sample-draw

Sample Flow

3.0 SCFH typical, with no inlet or exhaust line

Flow to Sensor

1.0 SCFH (nominal)

Maximum Recommended Inlet/
Exhaust Line Length for
1/4” O.D. x 1/8” I.D. Tubing

Inlet

Exhaust

100 feet

0 feet

50 feet

50 feet

Maximum Recommended Inlet/
Exhaust Line Length for
1/4” O.D. x 0.170” I.D. Tubing

Inlet

Exhaust

100 feet

100 feet

Response Time

90% in 30 seconds

Accuracy

± 5% of reading or ± 5 ppm CO (whichever is greater)