RKI Instruments 35-3001-04 User Manual

Page 8

Advertising
background image

4 • 35-3001-04 Carbon Monoxide Sample-Draw Detector

Main Circuit Board

The main circuit board includes the detector/amp terminal strip, interconnect terminal strip, oxygen
sensor terminal strip, pump connector, and flowmeter circuit board connector (see Figure 1).

Detector/Amp Terminal Strip

The detector/amp terminal strip is the upper twelve-point terminal strip in the bottom right corner
of the main circuit board. Use the detector/amp terminal strip to connect the CO amplifier to the
main circuit board.

NOTE: The CO amplifier is factory-wired to the main circuit board. See the “Installation” section

on page 7 for all wiring procedures related to the sample-draw detector.

Interconnect Terminal Strip

The interconnect terminal strip is the lower twelve-point terminal strip in the bottom right corner of
the main circuit board. Use the interconnect terminal strip to connect the sample-draw detector to a
controller.

Oxygen Sensor Terminal Strip

The oxygen sensor terminal strip is a two-point terminal strip in the upper right corner of the main
circuit board. This terminal strip is not used in this version of the sample-draw detector.

Pump Connector

The pump connector is the two-point connector below the oxygen sensor terminal strip in the upper
right corner of the main circuit board. Use the pump connector to connect the pump to the main
circuit board.

NOTE: The pump is factory-wired to the main circuit board. See “Installation” on page 7 for all

wiring procedures related to the sample-draw detector.

Flowmeter Circuit Board Connector

The flowmeter circuit board connector is a six-position connector in the upper left corner of the
main circuit board. Use the flowmeter circuit board connector to connect the flowmeter circuit
board to the main circuit board.

NOTE: The flowmeter circuit board is factory wired to the main circuit board. See “Installation”

on page 7 for all wiring procedures related to the sample-draw detector.

Flowmeter Circuit Board

The flowmeter circuit board is mounted to the left side of the main circuit board using standoffs. It
includes the flowmeter, sensor flow control valve, status LEDs, pressure switch, and pump reset
switch.

Flowmeter

The flowmeter is mounted to the right side of the flowmeter circuit board. You can see it through
the window in the door. A ball in the flowmeter column indicates the flow rate to the sensor. The
flowmeter measures the flow in the range 0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The
optimum flow rate is 1.0 SCFH.

Sensor Flow Control Valve

The sensor flow control valve is mounted to the flowmeter circuit board above the flowmeter. The
sensor flow control valve adjusts the flow rate to the sensor. Turn the valve’s knob

Advertising