Maintenance • 32

Calibration

Calibration requires you to perform both a zero adjustment and a span adjustment while in
Maintenance Mode 2. Calibration should be done on a periodic basis, or when you install
a new sensor or notice unusual gas readings. RKI Instruments Inc. recommends you
calibrate the Model GD-K77D every six months.

NOTE:

If the GD-K77D is installed in a clean room environment, the calibration frequency may be
extended to yearly.

Adjusting the Zero Setting

1.

Turn on the power to the Model GD-K77D. See “Start Up” for more information.

2.

Activate Maintenance Mode 1 by flipping up the control cover, then pressing and
holding the Mode switch for three (3) seconds until 1-1 appears on the display and the
SKIP lamp flashes.

3.

Use the FLOW ADJ switch to indicate 1-3 on the display.

4.

Press the TEST/SET switch. The display will show “ADJ.”

5.

Press and hold the TEST/SET switch until 2-1 appears on the display. The unit is now
in Maintenance Mode 2.

6.

Use the FLOW ADJ switch to show 2-2 on the display.