The ramp control – Yamato Scientific Vacuum controller Vacuum controller User Manual

Page 17

Advertising
background image

15

The Ramp Control

Display After Operation

Key

Explanation

1

•

Turn the power on.

2

•

After turning the power on, the fixed

value control set mode is
automatically selected. The previous
pressure set point appears blinking
on the indicator.

•

Set at 133hPa in the example on the

left.

•

The fixed value control lamp appears

blinking on the indicator.

•

The set mode lamp is on, indicating

that the set mode is ready for
operation.

3

OPERATION

MENU

•

Press OPERATION MENU to select

the ramp control mode. Then the
previous

「Ramp Level」 and 「Ramp

Time

」 appears alternately blinking

on the indicator.

•

The ramp control lamp blinks on the

operation mode indicator.

•

The set mode lamp is on, indicating

that the set mode is ready for
operation.

•

Refer to

「Editing the Set point of the

Ramp Control

」 (p.20),when you

change

「Ramp Level」 and 「Ramp

Time

」.

4

ENTER

•

Press ENTER and

「Ramp Level」

and

「Ramp Time」 are decided.

Then the ramp control starts. The
present pressure measurement
appears on the indicator.

•

The ramp control lamp is on, but stop

blinking.

•

The set mode ends and the set lamp

is off, The solenoid valve opens and
the solenoid valve lamp is on.

•

Until the pressure is reached to the

aimed set point, the solenoid valve
for control opens or closes according
to the ramp time. The ramp control
continues.