Shutdown, Restart – Yamato Scientific Vacuum controller Vacuum controller User Manual
Page 19

17
Shutdown
Display After Operation
Key
Explanation
1
•
The fixed value control is running.
2
ENTER
•
Press ENTER while operation runs,
and an operation stops at that point.
•
Present pressre measurement and
「StoP」appear alternately blinking
on the indicator.
•
All operation mode lamp is off and the
leak lamp is on, indicating that it is
opening.
•
The solenoid valve for control closes
and the leak valve opens.
Restart
Display After Operation
Key
Explanation
1
•
Stop
2
ENTER
•
When operation stopped, press
ENTER again. Returned to the
previous mode, which was set just
before shutdown, the operation
restarts in order to achieve the
previous set point. The present
pressure measurement appears. The
operation mode lamp is on, indicating
that the operation which was stopped
is now running as before.
•
The leak valve closes and the
solenoid valve for control opens at
this stage.