The free mode – Yamato Scientific Vacuum controller Vacuum controller User Manual

Page 18

Advertising
background image

16

The Free Mode

Display After Operation

Key

Explanation

1

•

Turn the power on.

2

•

After turning the power on, the fixed

value control set mode is
automatically selected. The previous
pressure set point appears blinking
on the indicator.

•

Set at 133hPa in the example on the

left.

•

The fixed value control lamp appears

blinking.

•

The set mode lamp is on, indicating

that the set mode is ready for
operation.

3

OPERATION

MENU

•

Press OPERATION MENU twice to

select the free mode.

「FrEE」appears

blinking on the indicator.

•

All operation mode lamp is off,

indicating that the control is
cancelled.

•

The set mode lamp is also off.

4

ENTER

•

Press ENTER and the present

pressure measurement appears.

•

The solenoid valve for control opens

so that the solenoid valve lamp is on.

•

The solenoid valve for control is kept

opening unless ENTER is pressed
again or the power is turned off.