28



NIRS Vision Tutorial – Operation

6

Routine Analysis

In Routine Analysis you use the models and equations you have built to analyze
unknown samples. All Routine Analysis parameters are brought together in the
Operations Method.

1.8

Setting up an Operations Method

Vision offers a wizard for setting up an operations Method.

Start Operations Method Wizard

1.

Click on Mode  Data Acquisition

2.

Click on Mode  Routine Analysis  OM Wizard.

3.

In the Select Operations Method dialog box, click on [New]

4.

In the Select Operation Method Standardization dialog box, set the

following:

Reference Standardization

enabled

Click on [OK]

5.

In the OM Wizard – OM Name dialog box, enter

OM Name “Tutorial”

Click on [Next]

Select a DCM

1.

In the OM Wizard – Select DCM dialog box, select

DCM

Diags

Click on [Timer]

2.

In the Configure Timer/Reference Frequency dialog box, set the

following:

In the Sample Frequency box:

Sample Collection Interval (sec.) 0
No delay between

enabled

In the Time Mode box:

Timed Acquisition

disabled

In the Reference Frequency box:

Every x Sample

enabled

Frequency x

10

Click on [OK]