3 functioning, 1 10-port valve, Port valve – Metrohm 872 Extension Module Liquid handling User Manual
Page 14: Figure 3, Port valve – positions

3.1 10-port valve
■■■■■■■■■■■■■■■■■■■■■■
6
■■■■■■■■
872 Extension Module – Liquid handling
3 Functioning
3.1
10-port valve
The 10-port valve is used for sample preparation. For this, the central posi-
tion (3-11) can be connected to the 10 other positions, one after the
other, by means of the selector (3-13).
The transfer tubing is connected to the central position. Liquids can be
aspirated into the transfer tubing by one of the 10 positions with the aid
of a Dosino and then transported to any other position by reversing the
flow.
The following illustration shows a 10-port valve with a possible utilization
of the positions:
1
2
3
4
5
6
7
8
9
10
11
12
13
Figure 3
10-port valve – positions
1
Position 1
Connector for the connection to e.g. the
injection valve in the extension module.
2
Position 2
Connector for the connection to e.g. the
injection valve in the IC instrument.
3
Position 3
Connector for the connection to e.g. the
mixing vessel.
4
Position 4
Connector for the connection to an auxiliary
solution.
5
Position 5
Connector for the connection to an auxiliary
solution.
6
Position 6
Connector for the connection to an auxiliary
solution.
7
Position 7
Connector for the connection to an auxiliary
solution.
8
Position 8
Connector for the connection to e.g. the
waste container.
9
Position 9
Connector for the connection to an auxiliary
solution.
10 Position 10
Connector for the connection to e.g. the ion
exchanger.