3.1 10-port valve

■■■■■■■■■■■■■■■■■■■■■■

6

■■■■■■■■

872 Extension Module – Liquid handling

3 Functioning

3.1

10-port valve

The 10-port valve is used for sample preparation. For this, the central posi-
tion (3-11) can be connected to the 10 other positions, one after the
other, by means of the selector (3-13).

The transfer tubing is connected to the central position. Liquids can be
aspirated into the transfer tubing by one of the 10 positions with the aid
of a Dosino and then transported to any other position by reversing the
flow.

The following illustration shows a 10-port valve with a possible utilization
of the positions:

1

2

3

4

5

6

7

8

9

10

11

12

13

Figure 3

10-port valve – positions

1

Position 1
Connector for the connection to e.g. the
injection valve in the extension module.

2

Position 2
Connector for the connection to e.g. the
injection valve in the IC instrument.

3

Position 3
Connector for the connection to e.g. the
mixing vessel.

4

Position 4
Connector for the connection to an auxiliary
solution.

5

Position 5
Connector for the connection to an auxiliary
solution.

6

Position 6
Connector for the connection to an auxiliary
solution.

7

Position 7
Connector for the connection to an auxiliary
solution.

8

Position 8
Connector for the connection to e.g. the
waste container.

9

Position 9
Connector for the connection to an auxiliary
solution.

10 Position 10

Connector for the connection to e.g. the ion
exchanger.

Advertising