4 10-port valve, 1 connecting the 10-port valve, Port valve – Metrohm 872 Extension Module Liquid handling User Manual

Page 36: Connecting the 10-port valve

Advertising
background image

5.4 10-port valve

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

872 Extension Module – Liquid handling

5.4

10-port valve

5.4.1

Connecting the 10-port valve

Connecting the 10-port valve

NOTE

Connect all capillaries with 6.2744.010 long PEEK pressure screws:

1 Connection to the transfer tubing

■

Connect the capillary with a long pressure screw to the central
connector of the 10-port valve.

■

Fix the other end of the capillary to the coupling on the transfer
tubing.

2 Connecting transfer tubing and Dosino

■

Fix the capillary to the coupling of the transfer tubing.

■

Connect the other end of the capillary to the Dosino (see manual
for the Dosino)
.

3 Other capillary connections

The other 10 connectors can be freely used for:

■

Connection to the injection valve in the extension module.

■

Connection to the injection valve in the IC instrument.

■

Connection to the mixing vessel.

■

Connection to different solutions.

■

Connection to an ion exchanger.

■

etc.