6 Troubleshooting

881 Compact IC pro – Cation

■■■■■■■■

71

Problem

Cause

Remedy

Very noisy baseline

High pressure pump – con-
taminated pump valves.

Clean pump valves (see Chapter 5.5.2, page
54)
.

Eluent – Leak in the eluent
path.

Check the eluent path

Eluent – Blockage in the
eluent path.

Check the eluent path

High pressure pump –
defective piston seals.

Replace (see Chapter 5.5.2, page 54) piston
seals .

Pulsation absorber not
connected. or defective.

Connect pulsation absorber (see Chapter 3.12,
page 33)
. or replace it.

Background conduc-
tivity too high

Incorrect eluent.

Change eluent (see Chapter 5.4.2.3, page 53).

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean the contact surfaces of the column chip
(with alcohol).

Column chip defective.

1. Save column configuration in MagIC Net™.
2. Notify Metrohm Service.

Individual peaks
greater than expec-
ted

Sample – carry-over of the
samples from previous
measurement.

Rinse system longer between two samples.

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path

Eluent – Blockage in the
eluent path.

Check the eluent path

Chromatograms
have poor resolution

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.10.4, page 69)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 45)
.

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Capillary connections –
dead volume in the system.

Check connections (see Chapter 3.5, page 14)
(use PEEK capillaries with an internal diameter
of 0.25 mm between the injection valve and
detector).

Guard column – diminished
performance.

Replace guard column (see Chapter 3.16,
page 42)
.