4.4 Maintenance and servicing

38

813 Compact Autosampler

To replace a pump tubing proceed as follows:

1 Remove old pump tubing

• Press contact pressure lever

11

on the tubing cartridge down

as far as it will go.

• Release tubing cartridge

11

from holding clamp

14

by press-

ing down snap-action lever

15

and remove from mounting

pin

10

at the 813 Compact Autosampler (see Fig. 1).

• Remove old pump tubing.

2 Insert new pump tubing

• Insert the new pump tubing

8

(6.1826.040) in the tubing

cartridge as shown in Fig. 9. The Stopper

32

must click into

the corresponding holder on the left-hand side of the tubing

cartridge.

• Place the tubing cartridge on mounting pin

10

and press

down on the right-hand side until snap-action lever

15

clicks

into position on holding clamp

14

. Take care that no kinks are

formed in the pump tubing.

3 Set contact pressure

• Press contact pressure lever

13

upwards until the solution

just starts to be drawn in. Then press contact pressure lever

upwards until it clicks once more to obtain optimal contact

pressure.

The 6.1826.0X0 pump tubing is made of PVC and must not be

used for rinsing with solutions which contain acetone. In such

cases, rinse with different pump tubing or a different pump.