3 Installation

850 Professional IC–Anion – MSM-HC – MCS – LP Gradient

■■■■■■■■

15

5 Installing the peristaltic pump

See Chapter 3.19.2, Page 59

Install pump tubings.

Connect aspirating capillaries for the regeneration solution and
the rinsing solution.

6 Installing the MSM-HC

Connect the MSM-HC capillary labeled with H2SO4 to the outlet
of the pump tubing for the regeneration solution.

Connect the MSM-HC capillary labeled with H2O to the outlet of
the pump tubing for the rinsing solution.

Guide the two MSM-HC capillaries labeled with Waste through an
instrument's capillary feed-through to the waste container and fix
them there.

7 Connecting the instrument

Connect the instrument to the PC (see Chapter 3.22.1, page
69)
.

Connect instrument to mains supply (see Chapter 3.22.2, page
69)
.

8 Initial start-up

See Chapter 4.1, Page 74

Switch on the PC and start MagIC Net.

Switch on and prepare the instrument.

Deaerating the high pressure pump.

Rinse the instrument without column(s).

9 Connecting the columns

Remove the 6.2744.040 coupling between the column inlet capil-
lary and the MSM-HC capillary labeled with Eluent.

Install the guard column (if used) (see "Connecting and rinsing
the guard column", page 71)
.

Install the separation column (see "Connecting and rinsing the
separation column", page 73)
.

10 Conditioning the instrument

See Chapter 4.1, Page 74

The instrument is now ready for measuring samples.