3 Installation

930 Compact IC Flex ChS/PP (2.930.1300)

■■■■■■■■

51

2 Installing the input of the separation column

There are 3 possibilities:

■

Attaching the bottom end of the separation column directly to
the guard column.
or

■

If the guard column is connected to the separation column using
a provided connection capillary, connect the bottom end of the
separation column to the guard column's output capillary using
the provided PEEK pressure screw (6.2744.070).
or

■

If no guard column is used (not recommended), connect the col-
umn input capillary to the input of the separation column using a
short pressure screw (6.2744.070).

3 Rinsing the separation column

■

Place a beaker under the separation column's outlet.

■

Start manual control in MagIC Net and select the high-pressure
pump: Manual

▶ Manual control ▶ Pump

– Flow: in accordance with column leaflet
– On

■

Rinse the separation column with eluent for approximately 10
minutes.

■

Use manual control in MagIC Net to stop the high-pressure pump
again: Off.

4 Removing the coupling

■

Remove the coupling (6.2744.040) from the column output capil-
lary.

5 Installing the output of the separation column

■

Fasten the column output capillary to the upper end of the sepa-
ration column using a PEEK pressure screw (6.2744.070).

6 Inserting the separation column

■

Insert the separation column with chip into the column holder
until you hear it snap in place.

The separation column is now detected by MagIC Net.