3 power connection, 4 leak sensor, 5 ambient conditions – Metrohm 896 Professional Detector – Conductivity User Manual
Page 31: Power connection, Leak sensor, Ambient conditions

■■■■■■■■■■■■■■■■■■■■■■
6 Technical specifications
2.896.0010 896 Professional Detector – Conductivity
■■■■■■■■
23
Temperature
stability
< 0.001 °C
Temperature
compensation
0…5 %/K adjustable, default 2.3 %/K
Heating up time
< 30 minutes (40 °C)
6.3
Power connection
Required voltage
100 - 240 V ± 10% (autosensing)
Required fre-
quency
50 - 60 Hz ± 3 Hz (autosensing)
Power consump-
tion
■
65 W for typical analysis application
■
25 W standby (conductivity detector to 40 °C)
Power supply unit
■
Up to 300 W maximum, electronically monitored
■
internal fuse 3.15 A
6.4
Leak sensor
Type
Electronic, no calibration necessary
6.5
Ambient conditions
Operation
Ambient tem-
perature
+5…+45 °C
Humidity
20…80 % relative humidity
Storage
Ambient tem-
perature
–20…+70 °C
Transport
Ambient tem-
perature
–40…+70 °C