5 reports, Reports – Metrohm 865 Dosimat plus User Manual
Page 58
6.2 Extended dosing (XDOS)
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
865 Dosimat plus
Selection
off
Default value
off
off
No safety limit.
Auto fill
Automatic filling of the cylinder after the dosing.
Selection
off | on
Default value
on
6.2.5
Reports
Menu
▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated result and additional specifica-
tions.
Selection
off | on
Default value
off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 34).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.
Selection
on | off
Default value
off