1.1 Instrument description

■■■■■■■■■■■■■■■■■■■■■■

2

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 1 (2.940.1510)

Leak sensor
The leak sensor detects leaking liquid that collects in the instrument's base
tray. Liquid that leaks in the instrument is routed to the base tray using
drainage tubing and detected there.

Column thermostat
The column thermostat regulates the temperature for the separation col-
umns and the eluent, thereby providing stable measuring conditions. The
interior of the column thermostat can be heated and cooled. There are
two column holders with chip readers in the column thermostat.

Eluent degasser
The eluent degasser removes gas bubbles and dissolved gases from the
eluent.

High-pressure pump
The intelligent and low-pulsation high-pressure pump pumps the eluent
through the system. It is equipped with a chip where its technical specifi-
cations and "life history" (operating hours, service data, etc.) are saved.

Inline filter
Inline filters protect the separation column reliably from potential contami-
nation from the eluent. The small filter pads with 2 µm pore size can be
replaced quickly and easily. They remove particles from the solutions, such
as bacteria and algae.

Pulsation absorber
The pulsation absorber protects the separation column from damage
caused by pressure fluctuations when switching the injection valve, and
reduces interfering pulsations during highly sensitive measurements.

Injection valve
The injection valve connects the eluent path to the sample path. By a
quick and precise switching of the valve a quantity of sample solution
defined by the size of the sample loop is injected and flushed to the sepa-
ration column with the eluent.

Suppressor
The suppressor consists of a suppressor drive, a rotor and, where applica-
ble, an adapter. The suppressor drive gives you the flexibility to use differ-
ent rotors according to the principle "one drive – many rotors". With
appropriate adapters, the rotor for the sample preparation module (SPM
Rotor) or suppressor rotors with different capacities and construction can
be easily exchanged. The rotors are not included in the instrument's scope
of delivery. A rotor suitable for the application must be ordered separately.