5.11 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

76

■■■■■■■■

882 Compact IC plus – Anion

■

Connect the inlet capillary for the rinsing solution (labeled with
rinsing solution) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040).

■

Rinse the first suppression unit with degassed ultra pure water for
about 15 minutes.

■

In the software, use the Step command to switch to the second
suppressor unit and rinse this for about 15 minutes.

■

In the software, use the Step command to switch to the third
suppressor unit and rinse this for about 15 minutes.

5 Connect the suppressor to the IC system

■

Reconnect the capillaries labeled regenerant and rinsing solution
to the IC system.

5.11.3.3

Cleaning the suppressor

In the following cases, it may be necessary to clean the suppressor:

■

Increased backpressure onto the connection tubings of the suppres-
sor..

■

Blockage of the suppressor which cannot be eliminated (solutions can
no longer be pumped through the suppressor).

■

Jamming of the suppressor which cannot be eliminated (suppressor
can no longer be switched over).

Cleaning the suppressor

Clean the suppressor as follows:

1 Disconnecting the suppressor from the IC system

■

Switch off the instrument.

■

Disconnect all capillaries of the suppressor from the IC system.

2 Dismantling the suppressor

■

Unscrew union nut (36-1) from the housing (36-4).

■

Pull the connecting piece (36-2) and the rotor (36-3) out of the
housing.
The connecting piece and the rotor normally stick to one another
– if this is not the case: Take a sharp object, insert it into the slot
of the housing (36-5) and pull out the rotor in this way.

■

Detach the connecting piece from the rotor.