6 Troubleshooting

882 Compact IC plus – Anion

■■■■■■■■

85

Problem

Cause

Remedy

Precision problems -
significant scatter-
ing of the measured
values

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 3.12.1, page 34)
.

Sample – rinsing volume
too low.

Increase rinsing time (see Chapter 5.9, page
69)
.

Injection valve – defective.

Request Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.12.4, page 80)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 48)
.

High pressure pump –
defective.

Request Metrohm Service.