3.5 Titration vessel for coulometric KF titration

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

851 Titrando

3.5.2

Coulometer cell – Standard setup

Filling the adsorber tube

Before setting up the coulometer cell the 6.1403.030 adsorber tube has
to be filled with 6.2811.000 molecular sieve. Proceed as follows:

6.1403.030

6.2811.000

1

2

3

4

Figure 13

Filling the adsorber tube

1 Insert a small cotton plug into the bottom of the adsorber tube. Do

not pack the cotton too tightly.

2 Fill the molecular sieve up to ¾ of the height.

3 Place a small cotton plug on the molecular sieve. Do not pack the

cotton too tightly.

4 Seal the adsorber tube with the appropriate cover.

NOTE

Note that the molecular sieve must be replaced at regular intervals.
Each time you refill the adsorber tube with molecular sieve, you can, for
example, write the date directly on the adsorber tube.