Metrohm 858 Professional Sample Processor User Manual

Page 34

Advertising
background image

3.8 Mounting the sample tube

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

858 Professional Sample Processor

3

Nut
4.766.4320 (or 6.2833.020 + 6.2744.080)

4

Ring wedge

5

Needle holder

6

Sample tube
6.2846.000 (zirconium oxide) or
6.1835.020 / 6.1835.040 / 6.1835.050
(made of PEEK)

This is how you mount needle and capillary:

1 Remove the PEEK pressure screw

Loosen and remove the nut (23-3) screwed onto the needle holder.

2 Insert the needle

■

Insert the needle from above part way into the opening of the
needle holder (23-5).

■

Pull the PEEK ring wedge (23-4) down over the needle from
above. The narrow side of the seal must face upwards.

3 Fasten the needle

■

Screw the nut (23-3) into the needle holder. Slight pressure will
be required from below to push the needle upward while doing
this.

■

Use your hand to screw the nut in the needle holder tight (do not
use any tools!).

4 Connect the capillary

■

Pull the 6.2744.010 PEEK pressure screw (23-2) over the end of
the capillary.

■

Manually screw tight the PEEK pressure screw with the capillary in
the nut of the needle holder. The capillary must be pressed tight
while doing so.