Figure 32, Inserting the tubing cartridge – Metrohm 858 Professional Sample Processor User Manual

Page 40

Advertising
background image

3.12 Installing the peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

858 Professional Sample Processor

Figure 32

Inserting the tubing cartridge

6 Connect the capillaries

■

Screw the respective capillaries tightly to the two tubing olives
with PEEK pressure screws (31-1).

Setting the flow rate
The contact pressure of the tubing cartridge must be adjusted in order to
regulate the flow rate. Proceed as follows:

1 Set the contact pressure

■

Press the contact pressure lever (31-5) all the way forward.

■

Switch on the drive of the peristaltic pump.

■

Then release the contact pressure lever gradually until the liquid is
suctioned for the first time.

■

Now release the contact pressure lever by two latch positions.

The contact pressure is now set optimally.