3.6 Connecting the eluent bottle

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

930 Compact IC Flex Oven/SeS (2.930.2400)

3.6

Connecting the eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing . The eluent aspiration tubing is installed on the input for the high-
pressure pump.

The tubing has to be be fed out of the instrument through a suitable
opening (see "Openings on the back panel", page 11) before the loose
end can be connected to the eluent bottle.

Accessories

For this step you need the following accessories:

These parts are part of the accessory kit Vario/Flex ONE (6.5000.010).

■

Eluent bottle (6.1608.070)

■

The eluent bottle cap GL 45 accessory set (6.1602.160)
This accessory set contains the bottle cap, an M6 tubing nipple, an M8
tubing nipple, two O-rings and an M6 and M8 threaded stopper.

■

The tubing adapter for aspiration filter accessory set (6.2744.210)
This accessory set contains a filter holder, a clamping screw and tubing
weighting.

■

An aspiration filter (6.2821.090)

Connecting the eluent aspiration tubing

1 Installing the eluent bottle cap (6.1602.160)

■

Start by pushing the M8 tubing nipple onto the loose end of the
eluent aspiration tubing, followed by the O-ring.

■

Push the loose end of the eluent aspiration tubing through the
M8 opening of the bottle cap and screw it on for the time being.

Figure 7

Installing the eluent bottle cap