3 cleaning the suppressor – Metrohm 930 Compact IC Flex Oven/SeS User Manual

Page 86

Advertising
background image

5.11 Suppressor

■■■■■■■■■■■■■■■■■■■■■■

78

■■■■■■■■

930 Compact IC Flex Oven/SeS (2.930.2400)

■

In the software, use the Step command to switch to the next
suppressor unit and regenerate it as described above.

■

As soon as all three suppressor units have been regenerated, dis-
connect the capillary labeled regenerant from the coupling.

3 Rinsing the suppressor

After regeneration, each of the three suppressor units must be rinsed
with degassed ultrapure water for about 15 minutes.

■

Connect the capillary labeled rinsing solution to the outlet of
the high-pressure pump using a coupling (6.2744.040).

■

Set the flow of the high-pressure pump to 0.5 mL/min in the soft-
ware.

■

Switch on the high-pressure pump.
If the pressure decreases during rinsing, slowly increase the flow
of the pump to a maximum of 2 mL/min. When doing so, make
sure that the pressure does not exceed 2 MPa!

■

After approx. 15 minutes, switch off the high-pressure pump.

■

In the software, use the Step command to switch to the next
suppressor unit and rinse it as described above.

■

As soon as all three suppressor units have been rinsed, disconnect
the capillary labeled rinsing solution from the coupling.

4 Connecting the suppressor to the IC system

■

Reconnect the capillaries of the suppressor labeled regenerant
and rinsing solution to the IC system.

■

Reconnect the outlet of the high-pressure pump to the IC system.

5.11.3.3

Cleaning the suppressor

In the following cases, it may be necessary to clean the suppressor:

■

Increased backpressure at the suppressor's connection tubing.

■

Blockage of the suppressor which cannot be eliminated (solutions can
no longer be pumped through the suppressor).

■

Jamming of the suppressor which cannot be eliminated (suppressor
can no longer be switched over).

Cleaning the suppressor

1 Disconnecting the suppressor from the IC system

■

Switch off the instrument.

■

Disconnect all capillaries of the suppressor from the IC system.