6 Handling and maintenance

800 Dosino

■■■■■■■■

31

6.3

GLP - Validation

Every drive and every dosing unit manufactured by the Metrohm Co. is
subjected to rigorous quality controls prior to shipment. Every dosing unit
is issued a quality certificate attesting conformance with the strict quality
criteria of the Metrohm Co. GLP( G ood L aboratory P ractice) requires,
among other things, periodic inspection of analytical measuring devices
with respect to precision and correctness on the basis of standard operat-
ing procedures S tandard O perating P rocedure, SOP ). This may also
include an inspection of dosing accuracy.

Recommended literature

■

Metrohm brochure "Quality management with Metrohm", detailed
information concerning the principles and procedural methods of
Good Laboratory Practice

■

Metrohm Applications Bulletin 283/1 "Validation of Metrohm burets"

The validation of burets is carried out by the Metrohm-Service with a spe-
cial software.

The Metrohm agents worldwide offer the possibility of on-site inspections
and certifications of dosing units and Dosinos with respect toaccuracy. It is
recommended that an accuracy inspection be performed when the dosing
cylinders and dosing pistons of a dosing unit are replaced.