7.1 Problems

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

800 Dosino

7 Troubleshooting

7.1

Problems

Problem

Cause

Remedy

Data of the dosing
unit cannot be read.

Data chip of the dosing
unit mechanically dam-
aged or impaired by chem-
icals.

■

Remove the dosing drive and set it up
again.

■

Clean the data chip and the contact surfa-
ces.

■

Have the data chip replaced by the
Metrohm Service Dept.

Dosing unit recog-
nized either not at
all or incorrectly.

The dosing drive was not
attached correctly.

■

Remove the dosing drive and set it up
again.

■

Check whether the dosing drive is correctly
seated.

■

Switch the control instrument off and on
again.

■

If necessary contact Metrohm Service Dept.

Dosino becomes hot

Dosing drive is overloaded.
Valve disc or dosing piston
is blocked.

■

Switch off the instrument immediately.

■

Disassemble the dosing unit (see the man-
ual for the dosing unit) and clean all of the
individual parts. Replace defective parts.

Dosino cannot be
actuated by the con-
trol device.

Connection between Dos-
ino and control device is
interrupted or án error has
occurred on the Dosino.

■

Check the cable connections.

■

Switch the control instrument off and on
again.

■

Check the dosing and filling rate.

■

If necessary contact Metrohm Service Dept.

No dosing takes
place at all

Tubing connections are
blocked or dosing unit is
not assembled correctly.

■

Check whether the dosing tip is blocked.

■

Check whether the dosing port is sealed off
with a stopper.

■

Check whether the VENT port is sealed off
with a stopper (vacuum in the supply bot-
tle!). The VENT port must be open for pres-
sure compensation.

■

Remove the dosing drive and check
whether the dosing piston is connected to
the dosing drive. The piston stopper must
be flush with the upper side of the hous-
ing.