4.4 Faults and malfunctions

871 Advanced Bioscan / Instructions for Use 8.871.1003

29

Signal "overload"

•

Wrong IC-Net parameter

•

Check settings

'Range'

•

Damaged working elec-
trode

•

Flow cell is not con-
nected properly

easuring poten-

•

Replace working electrode

•

Check cable connections
(REF: black, WE: red,
AUX: blue)

•

Optimize measuring po-
tential

•

Wrong m
tial

Considerable
press

•

Leak in system

•

Check connections and

ure drop

make leakproof

Con
pre

siderable

ssure rise

•

Contamination of the
filter in the 6.2821.100

•

Filter unit PEEK

•

Change of column pack-
ing by injection of con-

•

Regenerate the column

taminated samples

as been

used

Replace the 6.2821.110
Filter

(see section 4.2.1) or re-
place column

Note

•

Mounting screws of flow
cell overtightened; 25
µm spacer h



:

s should always

microfiltered.

Sample
be

•

Check screws of flow cell;
use 50
µm spacer

Chromatograms
with poor resolu-
tion, change in the
retention times

•

Deterioration in separa-
tion efficiency of the IC
column

•

Old Eluent

•

Ionic strength or pH of
the sample differs
strongly from eluent

•

Regenerate the column
(see section 4.2.1) or re-
place column

•

Replace eluent

•

Dilute sample or change
pH of sample

Extreme peak
broadening,
splitting (double
peaks)

•

Dead volume at the col-
umn ends

•

Dead volume in IC sys-
tem

•

Replace column

•

Check connections

No signal from
detector

•

No mains current

•

871 Advanced Bioscan
not switched on

•

Faulty fuse

•

Measuring cell not con-
nected

•

Connection to IC Net is
not working properly

•

Check mains supply and
voltage (see Section 2.2)

•

Switch on the 871 at the
rear panel

•

Replace fuse (see section
2.2.1
)

•

Check connection

•

Check connection