Metrohm 930 Compact IC Flex Oven/SeS/PP/Deg User Manual

Page 60

Advertising
background image

3.19 Initial start-up

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)

MagIC Net detects the instrument and all of its modules.

3 Starting equilibration

■

Start the equilibration in MagIC Net: Workplace

▶ Run ▶ Equi-

libration

▶ Start HW.

4 Deaerating the high-pressure pump

■

Push the end of the purging needle (6.2816.040) over the end of
the deaerating capillary on the purge valve.

■

Insert the syringe (6.2816.020) in the Luer connector of the purg-
ing needle.

■

Open the purge valve using the rotary knob (approx. ½ turn).

■

Switch on the high-pressure pump in MagIC Net.

■

Use the syringe to aspirate eluent until there are no more air bub-
bles in the aspirated eluent.

■

Switch off the high-pressure pump in MagIC Net.

■

Seal the purge valve using the rotary knob.

■

Remove the syringe from the purging needle.

■

Pull the purging needle out of the deaerating capillary.

Advertising