Metrohm 930 Compact IC Flex Oven/SeS/PP/Deg User Manual
Page 60

3.19 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)
MagIC Net detects the instrument and all of its modules.
3 Starting equilibration
■
Start the equilibration in MagIC Net: Workplace
▶ Run ▶ Equi-
libration
▶ Start HW.
4 Deaerating the high-pressure pump
■
Push the end of the purging needle (6.2816.040) over the end of
the deaerating capillary on the purge valve.
■
Insert the syringe (6.2816.020) in the Luer connector of the purg-
ing needle.
■
Open the purge valve using the rotary knob (approx. ½ turn).
■
Switch on the high-pressure pump in MagIC Net.
■
Use the syringe to aspirate eluent until there are no more air bub-
bles in the aspirated eluent.
■
Switch off the high-pressure pump in MagIC Net.
■
Seal the purge valve using the rotary knob.
■
Remove the syringe from the purging needle.
■
Pull the purging needle out of the deaerating capillary.