Application Bulletin 356

Installation instruction for «VoltIC pro I»

Page 12 of 18

For the data transfer from the VA Computrace software to

MagIC Net

an RS 232 device must be added to the device

list as follows:

Go to the Configuration window in MagIC Net

and select

«New» from the «Edit» button in the Device window. Select

the «RS 232 Device» from the «Miscellaneous» list:

Rename the device «797 RS232 Printer». Enter «1» in the

field «serial number».

Make the following settings under «RS 232»:

(Attention: the second newly generated COM port by the

null modem cable has to be selected as «COM Port», e.g.

COM 15 !)

Enter «stopp» in the field «Terminator for receive».

In the window method, import the method for VoltIC pro I

from the installation CD: Go to File

→ method manager and

choose your method group. Afterwards, click on edit

import and choose the pathway on the installation CD

(MagICnet\examples\methods\VoltICpro.imet).

The method carries out a parallel determination of cations

and

anions.

Immediately

after

starting

the

anion

determination, the cation determination (of the same

sample) is initialized. Samples could be taken from the

same vial, or from two vials with one of them containing the

acidified sample solution for cation determination. Both

measurements are recorded in the same data file. In order

to measure samples from two different vial please select

method «VoltICpro_mod.imet» from the same folder.

In the window method, open the VoltICpro method and

assign the following equipment: Assign the column