Metrohm 750 Autosampler User Manual

Page 15

Advertising
background image

2 Installation

750 Autosampler

12

Connect one end of the 6.2115.070 cable (option) to the

connections "EXTERNAL INJECT INPUT" and "GND" on the

connector plug inserted in the contact closure strip 10

10.

Connect the other end of the 6.2115.070 cable to the con-

nections "Pos. Fill" on the terminal block 38

38 (valve A) or 46

46

(valve B) of the 733 IC Separation Center so that COM and

GND are connected.

2 Electrical connection 750 – 709 (optional)

Connect jumper P12 to the control interface of the 709 IC

Pump in such a way that the contacts at outputs 3 and 4 will

be closed when the pump has stopped (see chap. 5.2 of the

709 Instructions for Use).

Use 6.2134.070 cable (option) to connect outputs 3 and 4 of

the control interface on the 709 IC Pump with connection

"REMOTE STOP INPUT" and "GND" on contact closure strip

10

10 of the 750 Autosampler.

Settings with 732 IC Detector as "Master"

The following settings are recommended for operating the IC system

with 732 IC Detector as "Master":

• 750 Autosampler:

"

Run Time

"

= 0.0 min

• 732 IC Detector:

operation with cycle program with

command "

valve A/B = fill

" at time

"

0.0 min

".

2.2.3

Connection to 762 with «IC Net»

When a modular IC system consisting of 732 IC Detector, 733 IC

Separation Center, 709 IC Pump, and 762 IC Interface is operated with

the PC software «IC Net» as "Master" the 750 Autosampler sample

change procedure will be started by a signal from the 733 IC Separa-

tion Center. The Autosampler must be connected to the 762 IC Inter-

face using the 6.2134.000 cable and to the 733 IC Separation Center

using the 6.2128.100 as shown in Fig. 7.

1 Electrical connection 750 – 762

Connect RS232 interface 11

11 at the 750 Autosampler to one of

the RS232 interfaces 14

14 or 18

18 at the 762 IC Interface using

the 6.2134.000 cable.

2 Electrical connection 750 – 733

Connect 6.2140.020 connector plug to contact closure strip 99

of the 750 Autosampler.

Connect one end of the 6.2128.100 cable to the connections

"LOAD NO", "LOAD COM", "INJECT NO" and "INJECT COM"

on the connector plug inserted in the contact closure strip 99.

Advertising