3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

■

Connect drainage tubings (see Chapter 3.6.5, page 27).

3 Connecting the eluent path

■

Equip the 6.1834.080 eluent aspiration tubing and connect with
the eluent bottle (see Chapter 3.8.1, page 32).

■

Connect the 6.1831.150 column inlet capillary and the capillary of
the MSM-HC labeled with Eluent with a 6.2744.040 coupling and
two 6.2744.070 short pressure screws.

■

Use a 6.2744.090 long pressure screw to connect the capillary of
the MSM-HC labeled with Detector to the input of the MCS (see
Chapter 3.19.2, page 65)
.

■

Connect the detector inlet capillary with a 6.2744.090 long pres-
sure screw to the output of the MCS (see Chapter 3.19.2, page
65)
.

4 Installing the upper peristaltic pump

(see Chapter 3.15.2, page 53)

■

Connect the 6.1803.020 aspirating capillary for regeneration solu-
tions with a 6.2744.034 tubing olive and a 6.2744.070 short
pressure screw to the aspiration end of the 6.1826.320 pump
tubing for the regeneration solution.

■

Place the pump tubing in a tubing cartridge.

■

Connect the 6.1803.020 aspirating capillary for rinsing solutions
with a 6.2744.034 tubing olive and a 6.2744.070 short pressure
screw to the aspiration end of the second 6.1826.320 pump tub-
ing for the rinsing solution.

■

Place the second pump tubing in the second tubing cartridge.

■

Place both tubing cartridges into the peristaltic pump.

5 Connecting the MSM-HC

(see Chapter 3.18.2, page 62)

■

Connect the capillary labeled with H2SO4 with a 6.2744.180
pump tubing connection and a 6.2744.070 short pressure screw
to the peristaltic pump on the outlet end of the pump tubing for
the regeneration solution.

■

Connect the capillary labeled with H20 with a 6.2744.180 pump
tubing connection and a 6.2744.070 short pressure screw to the
peristaltic pump on the outlet end of the pump tubing for the
rinsing solution.

■

Guide the two capillaries labeled with Waste into a waste con-
tainer and fasten it there.