14 sample preparation module (spm), 1 fields of application, 2 connectors – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual

Page 56: Fields of application, Connectors, See chapter 3.14

Advertising
background image

3.14 Sample preparation module (SPM)

■■■■■■■■■■■■■■■■■■■■■■

46

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

4 Guide the other end of the connection capillary out of the instrument

through a capillary feed-through.

Caution

If the sample degasser is not used, the input and output must be
sealed with the 6.2744.220 thread stoppers.

3.14

Sample preparation module (SPM)

3.14.1

Fields of application

The SPM consists of 3 neutralization units in total, which are used, regen-
erated and rinsed in rotation for sample preparation.

It can be used for the following applications (amongst others):

for cation exchange before feeding the sample (see "Cation
exchange", page 48)
.

for the neutralization of a sample (see "Neutralization", page 49).

3.14.2

Connectors

Caution

To protect the SPM against foreign particles or bacterial growth, a
pump tubing connection with filter (6.2744.180) must be mounted
between the peristaltic pump and the inlet capillaries of the SPM.

The three inlets and outlets of the neutralization units numbered 1..3 on
the SPM each have 2 fixed mounted PTFE capillaries (see Figure 24, page
47)
.

Advertising