37
• Adapt live display
If in the live display the measured value is not shown open with a right
click into the sub window Live display the dialogue window Proper-
ties – Live display. Change the following parameters:
Command
type
MEAS
Measured value display active
Measured value 1 Measured
value
• Carry out adjustment
Carry out the voltage adjustment as described in Section 5.2.2.
Titrino
Connect the Spectrosense to your Titrino (see Section 2.1) and turn it on.
• Select Mode
Press the <MODE> key and then press several times the arrow key
[Î] until
MEAS
appears. Confirm this selection with <ENTER>. In the
following dialogue line choose, also by pressing the key [Î] several
times, the
measured quantity U
and confirm again with <ENTER>.
• Start measurement
With the <MEAS/HOLD> key you can start the measurement. The
measured value is now shown on the display. To stop the measure-
ment press once more the <MEAS/HOLD> key.
• Carry out adjustment
Carry out the voltage adjustment as described in Section 5.2.2.