5 Appendix

Spectrosense, Instructions for Use

37

• Adapt live display

If in the live display the measured value is not shown open with a right
click into the sub window Live display the dialogue window Proper-
ties – Live display
. Change the following parameters:
Command

type

MEAS

Measured value display active
Measured value 1 Measured

value

• Carry out adjustment

Carry out the voltage adjustment as described in Section 5.2.2.

Titrino

Connect the Spectrosense to your Titrino (see Section 2.1) and turn it on.

• Select Mode

Press the <MODE> key and then press several times the arrow key
[Î] until

MEAS

appears. Confirm this selection with <ENTER>. In the

following dialogue line choose, also by pressing the key [Î] several
times, the

measured quantity U

and confirm again with <ENTER>.

• Start measurement

With the <MEAS/HOLD> key you can start the measurement. The
measured value is now shown on the display. To stop the measure-
ment press once more the <MEAS/HOLD> key.

• Carry out adjustment

Carry out the voltage adjustment as described in Section 5.2.2.