5 inserting the sample holder, Inserting the sample holder, Figure 17 – Metrohm 909 UV Digester User Manual
Page 25: Guiding the sample holder into the wet end
Advertising

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
909 UV Digester
■■■■■■■■
17
Note
Always insert the red plug into the red socket. This is the only way
that shielding against electrical interference can be ensured.
3.5
Inserting the sample holder
The sample holder has to be fully filled to ensure uniform temperature dis-
tribution in the samples (see "Preparing samples", page 26).
Proceed as follows:
1 Guide the sample holder on the mounting bracket into the wet end
so that the slot between sample positions 1 and 12 ends up located
at the front at the lamp holder (2-4).
12
1
Figure 17
Guiding the sample holder into the wet end
Ensure that the cylindrical guide on the bottom of the sample hold-
er's mounting bracket clicks in place in the holder (2-5).