Metrohm 909 UV Digester User Manual

Page 30

Advertising
background image

4.2 Dialog window overview

■■■■■■■■■■■■■■■■■■■■■■

22

■■■■■■■■

909 UV Digester

You can open the Lamp setup dialog by pressing and holding the [SET]
key while switching on the 909 UV Digester.

You have to edit the following parameters if you have inserted a new UV
lamp:

Reset the counters back to zero for the cumulative operating hours and
number of times the lamp has been switched on (Reset Counters).

Configure the number of operating hours for the new UV lamp until
the next replacement (Replace after).

Replace after

Maximum number of operating hours for the UV lamp until the next
replacement.

Input range

0 - 1000 h (Increment: 1)

Default value

750 h
We recommend replacing the UV lamp after 750 h
of operating time.

Reset counters

For resetting both counters for cumulative operating hours and the num-
ber of times the UV lamp has been switched on (ignitions) to 0.

Confirming with [OK] triggers the following confirmation prompt:

Selection

[OK] | [QUIT]

[OK]
Both counters are reset to 0.
[QUIT]
Both counters remain at their current values.

Caution

The counters cannot be reset during the ongoing service life of a UV
lamp currently in use.