Metrohm 850 Professional IC Cation - Prep 2 User Manual

Page 70

Advertising
background image

2.20 Separation column

■■■■■■■■■■■■■■■■■■■■■■

60

■■■■■■■■

850 Professional IC – Cation – Prep 2

Connecting and rinsing the separation column

1 Connecting the separation column

Caution

When using the columns, always ensure that these are correctly
inserted corresponding to the flow direction indicated.

■

Remove stoppers from the separation column.

■

Connect the lower end of the separation column with a
6.2744.070 short PEEK pressure screw to the outlet capillary of
the guard column (if used), or to the column input capillary .

2 Rinsing the separation column

■

Place beaker under the outlet end of the separation column.

■

Set the flow rate of the high pressure pump to the value suitable
for the separation column selected.

■

Start the high pressure pump and rinse the separation column
approx. 10 minutes with eluent.

■

Switch off the high pressure pump again.

3 Mounting the separation column

■

Hang separation column with chip into the column holder.

Note

The iColumns are equipped with a chip on which their operating data is
saved. The chip has to be hooked into the chip holder provided for this
so that the column recognition can function.

Advertising