Metrohm 791 VA Detector User Manual
Page 18

2 Installation
791 VA Detector
12
•
Screw nipple 36 onto screw nipple 33 to fix outlet capillary 37
is position.
4 Discharge
•
Pull outlet capillary 38 which is already mounted on cell body
32 through one of the openings in the rear panel of the 733 IC
Separation Center so that it emerges outside the instrument.
•
Insert outlet capillary 38 in a sufficiently large waste container
and fix it in position.
28
29
30
31
32
31
33
34
35
36
37
27
31
35
36
38
41
39
31
40
Fig. 5: ELCD cell assembly
27
ELCD cell (6.5303.030)
35
PTFE gasket (6.2704.020)
28
Working electrode
36
Screw nipple (4.420.2580)
29
Screw nipple (6.2730.000)
37
PEEK outlet capillary from
1.733.0X10 Detector block
30
Sealing ring (E.301.0028)
38
PTFE outlet capillary (6.1803.000)
31
PTFE gasket (6.2704.010)
39
Reference electrode (6.0727.000)
32
Cell body
40
Ag/AgCl Reference system
33
Screw nipple (6.2731.030)
41
Electrolyte vessel
34
Sealing ring (E.301.0004)