6 check of noise from current amplifier – Metrohm 791 VA Detector User Manual

Page 46

Advertising
background image

4 Maintenance – Malfunctions

791 VA Detector

40

4.4.6

Check of noise from current amplifier

1 Connect dummy cell.

2 Set digital switch 11 U

pol

to 0 V.

0

0

0

+

3 Set adjusting knob 8 to 1 nA.

4

5

1

10

50

1

5

0,1

0,5

100
10

0,5

0,1

4 Set commutator 3 "meas" to "on" position

(depressed position).

meas

5 Set commutators 4

7 to "odd" position

(non-depressed position).

6 Set selector 21 on the rear panel to 0

1 V.

0...10 mV

0...1 V

7 Allow the recording instrument to record for a

longer period of time (first adjust chart recorder

to middle of the scale). Typical deflections

caused by noise are about 10 mVpp or less (the

656 housing must be earthed).