Metrohm 901 Titrando User Manual

Page 11

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

901 Titrando

■■■■■■■■

3

1.3

Titration modes – Measuring modes – Dosing com-
mands

The 901 Titrando supports the following titration modes, measuring
modes and dosing commands:

■

SET
Endpoint titration at one or two specified endpoints.
Measuring modes:

– pH (pH measurement)
– U (potentiometric voltage measurement)
– Ipol (voltametric measurement with selectable polarization cur-

rent)

– Upol (amperometric measurement with selectable polarization

voltage)

■

KFT
Volumetric water content determination according to Karl Fischer.
Measuring modes:

– Ipol (voltametric measurement with selectable polarization cur-

rent)

– Upol (amperometric measurement with selectable polarization

voltage)

■

MEAS
The following measuring modes can be selected for measurements:

– pH (pH measurement)
– U (potentiometric voltage measurement)
– Ipol (voltametric measurement with selectable polarization cur-

rent)

– Upol (amperometric measurement with selectable polarization

voltage)

– T (temperature measurement)

■

CAL
Electrode calibration.
Measuring mode:

– pH (calibration of pH electrodes)

■

ELT
Electrode test for pH electrodes.
This mode is listed separately only in tiamo™. In Touch Control, the
electrode test is a component part of the CAL calibration mode.