Metrohm 758 KFD Titrino User Manual

Page 113

Advertising
background image

3.2 Remote control commands

758 KFD Titrino

109

3.2.2.52. MemoryCard.ChangeDir $G

MemoryCard.ChangeDir.Name

up to 10 ASCII characters

MemoryCard.ChangeDir.Checksum $G

MemoryCard.ChangeDir.Checksum.Value read

only

Changing the current directory. The action is carried out if "$G" is transmitted

to the corresponding node.

Do not use blank characters before and after name!

The checksum characterizes the content of the directory.

3.2.2.53. MemoryCard.CreateDir $G

MemoryCard.CreateDir.Name

up to 10 ASCII characters

Delete directory. The action is carried out if "$G" is transmitted to the corre-

sponding node.

Do not use blank characters before and after name!

3.2.2.54. MemoryCard.DelDir $G

MemoryCard.DelDir.Name

up to 10 ASCII characters

Delete directory. The action is carried out if "$G" is transmitted to the corre-

sponding node.

Do not use blank characters before and after name!

3.2.2.55. MemoryCard.Backup $G

MemoryCard.Backup.Name

up to 10 ASCII characters

Backup of the internal memory onto the card. The action is carried out if "$G"

is transmitted to the corresponding node.

Do not use blank characters before and after name!

3.2.2.56. MemoryCard.Reload $G

MemoryCard.Reload.Name

up to 10 ASCII characters

Reload a backup from the memory card into the internal memory. The action

is carried out if "$G" is transmitted to the corresponding node.

Do not use blank characters before and after name!

3.2.2.57. MemoryCard.Format $G

MemoryCard.CardLabel.Name

up to 8 ASCII characters

Format the memory card. The action is carried out if "$G" is transmitted to the

corresponding node.

Do not use blank characters before and after name!

3.2.2.58. MemoryCard.FreeMem read

only

Free memory on the card. $Q sends number of free bytes, e.g. "4928".

3.2.2.59. MemoryCard.BatteryChange $G

MemoryCard.BatteryChange.Date YYYY-MM-DD

Date for changing battery. The date will be set with "$G".

Advertising