3.2 Remote control commands

758 KFD Titrino

124

3.2.2.126. Assembly.Outputs.AutoEOD ON, OFF

Assembly.Outputs.SetLines $G

Assembly.Outputs.SetLines.L0

active, inactive, pulse, OFF

up to .L13

Assembly.Outputs.ResetLines $G

Setting the I/O output lines.

.AutoEOD: The automatic output of the EOD (End of Determination) at the

end of the determination can be switched off. Thus, for example,

in conjunction with a Titrino several determinations can be per-

formed in the same beaker. Before AutoEOD is switched on, line 3

must be set to "OFF".

.SetLines: With $G, all lines are set.

.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-

tive" means resetting of the signal, "pulse" means output of a

pulse of app. 150 ms, "OFF" means the line is not operated, see

also page 174.

Warnings:
• If you have "AutoEOD" to "ON", an active line 3 is set to "inac-

tive" by the EOD pulse.

• L6 is the line of the activate pulse. An active line 6 is set to "in-

active" by the activate pulse.

• L5 is the error line. It is continuously controlled by the Titrino

program and can therefore not be set freely.

Line assignments in Titrino program:

L0

Ready, inactive state

L1 Conditioning

OK

L2

Titration in progress

L3

EOD (End Of Determination)

L4

Can be set in TIP

L5 Error

L6

Activate pulse + can be set in TIP

L7

Buret volume pulses

L8–13 ---

.ResetLines: Lines are set to the inactive status (= high).

3.2.2.127. Assembly.Stirrer ON, OFF

Switching stirrer ON/OFF.

3.2.2.128. Setup.Comport

1, 2, 1&2

Selects the Titrino COM for the output of automatic info:

&Setup.Keycode

&Setup.Trace

&Setup.SendMeas

&Setup.AutoInfo