3.2 Remote control commands

784 KFP Titrino

64

&Setup

Object

Description

Input range

Reference

& Root

:

├

Setup

Settings for the operating mode

├

.Comport

Output of automatic info

1,2,1&2

3.2.2.82.

├

.Keycode

Send key code

ON, OFF

3.2.2.83.

├

.Tree

Sending format of path info

│ ├

.Short

Short format of path

ON, OFF

3.2.2.84.

│ ├

.ChangedOnly

Paths of modified nodes only

ON, OFF

ditto

│

├

.Trace

Message on changed values

ON, OFF

3.2.2.85.

│

├

.Lock

Lock key functions

│ ├

.Keyboard

Lock all keyboard keys

ON, OFF

3.2.2.86.

│ ├

.Config

Lock <CONFIG> key

ON, OFF

ditto

│ ├

.Parameter

Lock <PARAM> key

ON, OFF

ditto

│ ├

.SmplData

Lock <SMPL DATA> key

ON, OFF

ditto

│ ├

.UserMeth

Lock functions

│ │ ├

.Recall

Lock “loading”

ON, OFF

ditto

│ │ ├

.Store

Lock “saving”

ON, OFF

ditto

│ │ ├

.Delete

Lock “deletion”

ON, OFF

ditto

│ └

.Display

Lock display function

ON, OFF

ditto

│

├

.Mode

Setting waiting intervals

│

├

.StartWait

Waiting time after start

ON, OFF

3.2.2.87.

│

├

.FinWait

Waiting time after run

ON, OFF

ditto

│

├

.SendMeas

Automatic sending of measured values

│

├

.SendStatus

Connect/disconnect sending

ON, OFF

3.2.2.88.

│

├

.Interval

Time interval

0.08...4...16200,

ditto

│

MPList

│

├

.Select

Selection

Assembly, Titrator

3.2.2.89.

│

├

.Assembly

From assembly

│ │ ├

.CyclNo

Cycle number

ON, OFF

3.2.2.90.

│ │ ├

.V

Volume

ON, OFF

ditto

│ │ ├

.Meas

Measured indicator voltage

ON, OFF

ditto

│ ├

.Titrator

From Titrator

│ │ ├

.CyclNo

Cycle number

ON, OFF

3.2.2.91.

│ │ ├

.V

Volume

ON, OFF

ditto

│ │ ├

.Meas

Measured indicator voltage

ON, OFF

ditto

│ │ ├

.dVdt

Volume drift dV/dt

ON, OFF

ditto

│ │ ├

.dMeasdt

Measured value drift

ON, OFF

ditto

│ │ ├

.dMeasdV

1st deviation of titration curve

ON, OFF

ditto

│