Metrohm 784 KFP Titrino User Manual

Page 87

Advertising
background image

3.2 Remote control commands

784 KFP Titrino

82

.MeasPt.X"165" Time of the MPList in s
.MeasPt.Y"3.654" Volume of the MPList in mL
.MeasPt.Z1"6.34" Measured value of the MPList, format depends on the

measured quantity

.MeasPt.Z2 without meaning
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP


3.2.2.72. Info.ActualInfo.Oven.HeatTime

read only

Info.ActualInfo.Oven.SampleTemp

read only

Info.ActualInfo.Oven.LowTemp

read only

Info.ActualInfo.Oven.HighTemp

read only

Info.ActualInfo.Oven.GasFlow

read only

Info.ActualInfo.Oven.UnitFlow

read only

$Q sends the current values from a connected KF Oven. If no Oven is con-
nected, the values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.
.GasFlow: Average gas flow during sample heating time.
.UnitFlow: Unit of gas flow.


3.2.2.73. Info.ActualInfo.Display.L1

up to 32 ASCII characters

Info.ActualInfo.Display.L8

up to 32 ASCII characters

Info.ActualInfo.Display.DelAll

$G

Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 3.2.2.86.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.86.
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.46) to refresh the

display.

$Q sends the contents of the corresponding display line.


3.2.2.74. Info.ActualInfo.Comport.Number

read only

$Q sends the comport number of the Titrino where the PC is connected.


3.2.2.75. Info.Assembly.CycleTime

read only

Info.Assembly.ExV

read only

Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.