5.12 Suppressor

■■■■■■■■■■■■■■■■■■■■■■

94

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 2 (2.940.1520)

Replacing parts of the suppressor

1 Disconnecting the suppressor from the IC system

Switch off the instrument.

Disconnect all capillaries of the suppressor from the IC system.

2 Dismantling the suppressor

Unscrew union nut (25-1) from the housing (25-4).

Pull the connecting piece (25-2) out of the housing together with
the the rotor (25-3).
If rotor A gets stuck in the housing, you can knock it loose as fol-
lows:
Put a pointed object into the slot in the housing and use it to
knock rotor A loose.

Detach the connecting piece from the rotor.

3 Cleaning the new rotor

Clean the surface of the new rotor (25-3) with ethanol using a
lint-free cloth.

4 Inserting the new rotor

CAUTION

Rotor A may be destroyed during start-up if not inserted correctly.

Insert the new rotor (25-3) into the housing (25-4) in such a way
that the tubing connections on the rear of rotor fit into the corre-
sponding recesses inside the housing and one of the three holes
of the rotor is visible from below in the slot of the housing (25-5).