Metrohm 850 Professional IC Anion MSM-HC - Gradient User Manual

Page 24

Advertising
background image

3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

850 Professional IC Anion MSM-HC HP-Gradient

■

Connect the 6.1803.020 aspirating capillary for rinsing solutions
with a 6.2744.034 tubing olive and a 6.2744.070 short pressure
screw to the aspiration end of the second 6.1826.320 pump tub-
ing for the rinsing solution.

■

Place the second pump tubing in the second tubing cartridge.

■

Place both tubing cartridges into the peristaltic pump.

6 Connecting the MSM-HC

(see Chapter 3.16.2, page 47)

■

Connect the capillary labeled with regenerant with a 6.2744.180
pump tubing connection and a 6.2744.070 short pressure screw
to the peristaltic pump on the outlet end of the pump tubing for
the regeneration solution.

■

Connect the capillary labeled with Rinsing solution with a
6.2744.180 pump tubing connection and a 6.2744.070 short
pressure screw to the peristaltic pump on the outlet end of the
pump tubing for the rinsing solution.

■

Guide the two capillaries labeled with Waste reg. and Waste rins.
into a waste container and fasten them there.

7 Connecting the instrument

■

Connect the instrument to the PC with a 6.2151.020 USB cable
(see Chapter 3.18.1, page 55).

■

Connect instrument to mains supply (see Chapter 3.18, page
55)
.

8 Initial start-up

(see Chapter 4.1, page 59)

■

Switch on the PC and start MagIC Net.

■

Switch on the instrument.

■

Deaerate the high pressure pump (see Chapter 3.10.2, page
36)
.

■

Set contact pressure of the peristaltic pump (see "Set flow rate",
page 54)
.

■

Rinse instrument without column.

9 Installing guard and separation column

■

Remove the 6.2744.040 coupling between the column inlet capil-
lary and the MSM-HC capillary labeled with in.