Metrohm 886 Prof.Thermostat/Reactor User Manual

Page 34

Advertising
background image

3.8 Installing accessories

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

1

2

3

4

1 Lowering the latches

Loosen the outer screws of both latches with the 6.2621.140 2.5
mm hexagon key turning them counterclockwise.

The latches open and unblock the insertion rails of the heating plate.

2 Insert reactor plate

Insert the reactor plate into the oven by fitting the two bolts at
the sides of the reaction plate into the insertion rails of the heat-
ing plate.

Slide the reactor plate into the oven as far as it will go. Take care
not to jam or pinch the capillaries.

3 Lift the latches

Tighten the outer screws of both latches with the 6.2621.140 hexa-
gon key 2.5 mm turning them clockwise.

The latches lift and fix the column holder on the heating plate.

4 Connect the capillaries to the system

In case the free capillary ends are not yet guided through the capil-
lary feed-throughs, guide them out of the oven.

Connect the loose end of the reactor inlet capillary (13-1) with
the outlet of the separation column.

Advertising