1 prepare instruments for diagnostic test, 2 perform display test – Metrohm 751 GPD Titrino User Manual

Page 197

Advertising
background image

4.2. Diagnosis

751 GPD Titrino

194

4.2.3 Equipment required:

Equipment required:

− voltage calibrator,

e.g. 1.642.0010 Metrohm pH Simulator

or 1.767.0010 Calibrated Reference for mV, pH, ê æS, °C

− highly insulated interconnection cable

6.2108.060

− resistor switch-box, class 0.1 % (or resistor 14.3 k, 0.1 %)
− cable 3.496.5070
− exchange units, if possible with different cylinder volumes (or 3.496.0070 dummy exchange unit)
− stop watch or watch with second hand
− digital or analogue voltmeter (if need be, connect a calibrated recorder)
− 2 connecting cables with 4 mm banana plugs
− test plug 3.496.8550 (necessary only if plug ‘Remote’ should be checked)
− test plug 3.496.8560 (necessary only if plug ‘RS 232’ should be checked)

1 Prepare instruments for diagnostic test

•

Power off.

•

Disconnect all external connections (cables at rear, except mains cable and keyboard).

•

Remove exchange unit.

•

Power on and immediately press and hold the <9> key until the powerup test pattern dis-

appears.

diagnose press key 0...9

2 Perform display test

•

Press <2>.

display test

•

Press <Enter>.

Characters for a visual check of the display are generated on the eight lines.

Test sequence:

a) The display is cleared and overwritten from the top left to the bottom right with a dot

pattern ( ²²²²²²²² ).

b) The display is cleared and overwritten from the top left to the bottom right with a dot

pattern( °°°°°°°°.

c) The display is continuously cleared and overwritten from the top left to the bottom

right with the complete character set (see page 195 / Fig. 1). At the same time with

moving display the LED’s „COND.“, „STATISTICS“ and „SILO“ are swtiches on and

off.

•

The test sequence can be held and then continued at any time by pressing <5>.

•

Block 2 is quit by pressing <Clear>.

diagnose press key 0...9