background image

4.2. Diagnosis

751 GPD Titrino

202

•

Press <Start>.

The internal dosing unit expells twice the volume of 3.5 ml.

•

Remove exchange unit.

•

Check spindle zero position, see Fig. 6.

0.2 - 0.6 mm

Fig. 6

Fig. 7

The spindle must be 0.2 - 0.6 mm below the edge of the sliding plate.

The bar of the stopcock coupling must be exactly parallel to the lateral edges of the

GPD Titrino, see Fig.7.

•

Reinsert Exchange Unit.

Titrino fills.

The display of before reappears.

•

Knob 'dV/dt' to right stop.

•

Press the <DOS> key (on instrument) until the piston rod reaches the top and at the same

time measure the time from start to end.

XXX X

********

cylinder empty !

mind selected language!

Spindle remains at maximum position. The transit time of the spindle is

20 s.

•

Measure spindle lifting (can be performed only if the 3.496.0070 Dummy Exchange Unit is

inserted or the locking switch (in right hole) is carefully operated with a screwdriver after

removal of the Exchange Unit).

From the start point, the spindle travels 80 mm. Instead of the spindle height, the ex-

pelled volume can be measured (corresponds to max. vol. of Exchange Unit used).

•

Press <FILL> (on instrument) and simultaneously measure the time until the GPD Titrino

is again in the 'ready' position.

Times for filling:

per stop cock cycle 1 s

for filling

20 s (Tolerance: 10 %)

The following generally holds: Spindle and stopcock must move at a constant speed

(noise!). In the filling setting, the stopcock coupling must position the lever of the

Exchange Unit correctly at the left stop (with virtually no play and without sticking).

•

Set potentiometer 'dV/dt' to left stop.

•

Press <DOS> (on instrument) at same time and use a stopwatch to measure the time for
1

/

10 of the cylinder volume to be expelled. The time should be ca. 76 ... 126 s.

•

Set potentiometer 'dV/dt' to right stop.

•

Press <FILL>.