4.13 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
940 Professional IC Vario ONE/ChS/PP/LPG (2.940.1350)
Connecting capillaries for rinsing solution (as an alternative
to STREAM)
Accessories
For this step you need the following accessories:
■
Aspiration capillary (6.1803.020)
■
2 × pressure screw, short (6.2744.070)
On the pump tubing's …
… connect this accessory
Input
Tighten the aspiration tubing (6.1803.020) to the tubing olive (6.2744.034)
using a pressure screw (6.2744.070).
Output
Tighten the regenerant suppressor capillary to the pump tubing connec-
tion with safety device (6.2744.180) using a pressure screw (6.2744.070).
Next step
■
Connect the loose end of the aspiration capillary to the bottle with the
rinsing solution.
Setting the contact pressure correctly
1
■
Fully loosen the contact pressure lever , i.e. press it all the way
down.
■
Activate the peristaltic pump drive in the software.
■
Raise the contact pressure lever one step at a time until liquid
flows.