4.13 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

940 Professional IC Vario ONE/ChS/PP/LPG (2.940.1350)

Connecting capillaries for rinsing solution (as an alternative
to STREAM)

Accessories

For this step you need the following accessories:

Aspiration capillary (6.1803.020)

2 × pressure screw, short (6.2744.070)

On the pump tubing's …

… connect this accessory

Input

Tighten the aspiration tubing (6.1803.020) to the tubing olive (6.2744.034)
using a pressure screw (6.2744.070).

Output

Tighten the regenerant suppressor capillary to the pump tubing connec-
tion with safety device (6.2744.180) using a pressure screw (6.2744.070).

Next step

Connect the loose end of the aspiration capillary to the bottle with the
rinsing solution.

Setting the contact pressure correctly

1

Fully loosen the contact pressure lever , i.e. press it all the way
down.

Activate the peristaltic pump drive in the software.

Raise the contact pressure lever one step at a time until liquid
flows.