7 How to ...?

VA Database 2.0

72

Send sample tables to the backup directory

1. If you want to observe the process of data transfer, click on

or

MAIN WINDOW / Window / Communication

to open the

VA

COMMUNICATION

window (the data transfer works also in the

background).

2. On the dialog page "SAMPLES" in the 746 VA Trace Analyzer

press the softkeys

< >> >

and

<Copy to RS Ifc.1>

.

The current sample table will be transferred as ASCII sample ta-

ble file (

*.smp

) to the predefined backup directory (see Default

directories, page 51).

Note: The transfer of sample table files can only be done via RS
Interface 1 at the 746 VA Trace Analyzer.

Send rack definition files to the backup directory

If a 695 Autosampler is connected to the 746 VA Trace Analyzer,

the rack data configuration files stored in the 695 Autosampler
can be saved first on a data card and then on the PC. Proceed

as follows:
1. If you want to observe the process of data transfer, click on

or

MAIN WINDOW / Window / Communication

to open the

VA

COMMUNICATION

window (the data transfer works also in the

background).

2. Press the softkeys

<Settings>

,

<695 Autosampler>

, and

<Rack

def>

on the dialog page "STAND" in the 746 VA Trace Ana-

lyzer.

3. Load the desired rack data by entering the

Rack name

and

pressing the softkey

<Load from 695>

.

4. Store the rack definition file to the data card inserted in the

746 VA Trace Analyzer by pressing the softkey

<Copy to Data

Cd.>

.

5. Select the desired rack definition file (

*.aut

) on the dialog

page "DATA CARD" and press the softkeys

<Copy to>

and

<RS232 Ifc.1>/<RS232 Ifc.2>

.

The current rack definition data will be transferred as binary rack
definition file (

*.aut

) to the predefined backup directory (see De-

fault directories, page 51).