887 Professional UV/VIS Detector

4 Start-up

The 887 Professional UV/VIS Detector is set up together wit other instru-
ments, e.g. an 850 Professional IC and the 886 Professional Thermostat/
Reactor.

Putting the 887 Professional UV/VIS Detector into opera-
tion

1 Start MagIC Net.

2 Switch the 887 Professional UV/VIS Detector on.

The 887 Professional UV/VIS Detector is recognized automatically by
MagIC Net.

Evaluating the intensity spectrum

Before you can evaluate the intensity of the lamps, the following precon-
ditions must be met:

■

The lenses and the flow path of the flow-through cell must be clean.

1 Rinsing the flow-through cell with ultrapure water

Rinse the flow-through cell with ultrapure water at a flow of 0.5
mL/min.

Ensure that no air bubbles remain in the flow-through cell.

2 Evaluating the intensity of the lamps

In MagIC Net™, adjust the following settings:

■

Go to program part Manual.

■

Click on the symbol of the 887 UV/VIS Detector.

■

On the UV lamp tab, switch on the UV Lamp.

■

On the VIS lamp tab, switch on the VIS lamp.

■

On the Detector tab, select the Intensity spectrum.
First click on [Reset baseline] and then click on [View].

The lamps intensity range is monitored an a new spectrum is recor-
ded.

If the intensity spectrum looks similar to the spectrum in Figure 8, then the
lamp is correctly adjusted.